Part 3 - keypad configuration, K-factor programming – Dynasonics TFXP Series Transit Time Ultrasonic Flow Meter User Manual

Page 68

background image


Rev. 5/09

-3.30-

TFXP

Example 2: To span the 0 to 2.5k Hz output from 0 GPM to +100
GPM, with 1.25k Hz being 50 GPM, set the FL 0H and FL 25KH
inputs as follows:

FL 0H = 0.0

FL 25KH = 100.0

K-Factor Programming

If the device receiving the pulse output is capable of K-factor
scaling, it is possible to use the rate pulse output in a totalizing
function. See K-Factors Explained in the Appendix.

Figure 3.11 — Rate Pulse Output Wiring

K-Factor
Programming