7: Processing Menu Functions

84

000-20000-300-02- 201110

Procedure

To perform reference monitoring:

1. Start an acquisition on the spectrometer that will monitor the lamp. Adjust the acquisition

parameters so that the spectrum is on scale.

2. Store a reference and dark spectrum for that acquisition.

3. Start an acquisition on the spectrometer that will measure the sample.

4. Store a dark spectrum for the sample acquisition.

5. Click on the sample acquisition graph,

Or,

In the Data Sources pane, select the processing node for the sample acquisition (No pre-
processor
, etc.). The reference monitoring action should enable.

6. Click

Or,

Select Processing | Preprocessing | Add Reference Monitor. The Configure Reference
Monitoring
dialog box appears:

7. Select the desired reference channel.

8. Select the Reference Type as follows: