Calibration combine calibration files, Calibration, Combine calibration files – Ocean Optics SpectraSuite User Manual
Page 102

7: Processing Menu Functions
90
000-20000-300-02- 201110
Fiber Diameter – If the light source is not inside the integrating sphere, you must enable
this option and enter a value in the field (in microns) equivalent to the diameter of the
opening of the integrating sphere that is capturing the measured light. There are 25,400
microns per inch. For example, for the FOIS-1, the value would be 9525 microns (equals
3/8 of an inch).
However, if a light source is inserted into the sphere, this value is ignored. If a cosine
corrector is used, define the light collection area as follows:
If this cosine corrector is used . . .
Enter this diameter (in um) . . .
CC-3 or CC-3-UV
3900
CC-3-DA
7140
SpectraSuite then calculates the collection area from the fiber diameter.
Calibration
You can calibrate your spectrometer in two ways:
Ocean Optics can calibrate the spectrometer for you and send you the resulting
calibration file, or
You can calibrate the spectrometer yourself using a light source (calibration lamp) with
known spectral output (microwatts per square centimeter per nanometer) using this menu
item.
Combine Calibration Files
This feature allows you to combine two spectrometer calibration files that you have created; one file
using a calibrated deuterium lamp source (UV) and the other using a calibrated tungsten lamp source
(VIS). This will result in a spectrometer calibration file that can be used across the entire spectral range.