If the signal you collect is saturating the spectrometer (intensity greater than the maximum value for the
spectrometer), you can decrease the light level on scale in scope mode by:

 Decreasing the integration time

 Attenuating the light going into the spectrometer

 Using a smaller diameter fiber

 Using a neutral density filter with the correct optical density

If the signal you collect has too little light, you can increase the light level on scale in scope mode by:

 Increasing the integration time

 Using a larger diameter fiber

 Removing any optical filters

Taking Reference and Dark Spectra

Reference and dark spectra must be stored before collecting experiment data. If you pass the cursor over
the No-preprocessor line in either the Data Sources or Data Views pane, a dialog box informs you
whether reference and/or dark spectra have been stored.