Ocean Optics SpectraSuite User Manual
Page 84
6: Spectrometer Menu Functions
72
000-20000-300-02- 201110
Many of the features listed for the HR4000 Spectrometer are also valid for other spectrometers. The table
below lists all spectrometer features. Check your spectrometer to see which features apply to your model.
Feature
Meaning
Analog In
Set the input voltage and the update interval. Analog input can be
accessed from the spectrometer’s accessory connector.
Analog Out
Analog output can be used for multiple applications such as single pixel
intensity analysis or programmable reference/dimmer to light sources.
Set the analog output on this screen.
Autonulling
Autonulling is an enhanced electrical dark correction. Set the autonulling
status and value on this screen.
Board Temperature
The HR and QE spectrometers contain a temperature sensor chip on the
printed circuit board. Use this screen to set the temperature parameters.
Channel Features
For the ADC1000-USB and spectrometers such as Jaz with more than
one channel. Enable each channel by checking the box.
Continuous Strobe
Set the continuous strobe delay. The Continuous Strobe signal is a
programmable frequency pulse-train.
Detector Linearity
Use this screen to change the linearity coefficients.
External Trigger Delay
Set the counts for the external hardware trigger.
GPIO
Set the General Purpose Input/Output values on this screen.