beautypg.com

Source scpi command reference – AMETEK SFA Series Ethernet Programming User Manual

Page 85

background image

SG Series Programming

SCPI Command Operation

M550129-03 Rev K

6-13

[:AMPLitude]

[:AMPLitude?]

:PROTection

:PROTection?

[:LEVel]

[:LEVel?]

:TRIPped?
:STATe?
:CLEar

:RAMP

:RAMP?

:ABORt

:ALL?

:HTRIGgered

:HTRIGgered?

:TRIGgered

:TRIGgered?

:TRIGgered

:TRIGgered?

:CLEar

:AMPLitude
:AMPLitude?

6.5.2

SOURCE SCPI Command Reference

The letter “C” in the “SCPI” column means that the command syntax is
SCPI compliant; an “N” in the “SCPI” column means that the command
syntax is not part of the SCPI definition.

Command

Description

SCPI

SOURce

Source subsystem.

C

:CURRent

Sets the output current in amps (default) or in milliamps.

C

:CURRent?

Returns the output current in amps or in milliamps

C

[:LEVel]

Sets the output current in amps (default) or in milliamps.

C

[:LEVel?]

Returns the output current in amps or in milliamps.

C

[:IMMediate]

Sets the output current in amps (default) or in milliamps.

C

[:IMMediate?]

Returns the output current in amps or in milliamps

C

[:AMPLitude]

Sets the output current in amps (default) or in milliamps.

C

[:AMPLitude?]

Returns the output current in amps or in milliamps

C

:LIMit

Sets an upper soft limit on the programmed output current
for the supply. The soft limit prevents the supply from being
inadvertently programmed above the soft limit, thus
providing a method for protecting the load against
damaging currents.

C

:LIMit?

Returns the upper soft limit on the programmed output
current for the supply.

C

[:AMPLitude]

Sets an upper soft limit on the programmed output current for
the supply.

C

[:AMPLitude?]

Returns the upper soft limit on the programmed output
current for the supply.

C