Analog program adjustment – AMETEK SFA Series Ethernet Programming User Manual

Page 145

background image

SG Series Programming

Calibration

M550129-03 Rev K

7-11

7.8

ANALOG PROGRAM ADJUSTMENT

This sections applies only to units with the Isolated Interface Adapter *(IIA)
option.

All calibration potentiometers can be adjusted through access holes in the top
cover of the SGI Series unit. It is not necessary to remove the top cover to
perform the calibration procedure. Please refer to Figure 7-1 when following
the adjustment procedures.