SCPI Command Operation

SG Series Programming

6-30

M550129-03 Rev K

Command

Description

SCPI

:GAIN?

Returns the value of the gain for current
measurements.

N

:OFFSet

Sets the value of the offset for current
measurements.

N

:OFFSet?

Returns the value of the offset for current
measurements.

N

:POINt <1|2>

Sets the current measurement calibration point
(1 or 2). The actual output current is measured with
an external meter.

N

:VOLTage

Calibrate Measure Voltage subsystem

N

:ADC?

Returns the integer value of the A/D for the voltage
measurement.

N

:CALCulate

Calculates the value of the gain and offset for voltage
measurements.

N

:GAIN

Sets the value of the gain for voltage measurements.

N

:GAIN?

Returns the value of the gain for voltage
measurements.

N

:OFFSet

Sets the value of the offset for voltage
measurements.

N

:OFFSet?

Returns the value of the offset for the voltage
measurements.

N

:POINt <1|2>

Sets the voltage measurement calibration point
(1 or 2). The actual output voltage is measured with
an external meter.

N

:MODel

Calibrate Model subsystem

N

:LASTCALDATE

(Ethernet only) Sets the date last calibrated;
format: MM DD YYYY (space after MM and DD
required)

N

:LASTCALDATE?

(Ethernet only) Returns the date last calibrated.

N

:NEXTCALDATE

(Ethernet only) Sets the date next calibration is
required;
format: MM DD YYYY (space after MM and DD
required)

N

:NEXTCALDATE?

(Ethernet only) Returns the date next calibration is
required.

N

:POWERON

Sets the default output enable condition at power on.
Input format:
“ON,INIT” enables the output at next power on
“OFF,INIT” disables the output at next power on
NOTE: Quotation marks are required in the
command string.

N

:POWERON?

Returns the status of the output enable condition at
power on

N

:RESET

Sets the default output enable condition when the
*RST command is issued.
Input format:
“ON,INIT” enables the default output
“OFF,INIT” disables the default output
NOTE: Quotation marks are required in the
command string.

N

:RESET?

Returns the status of the default output enable
condition when the *RST command is issued

N