Calibration
SG Series Programming
7-8
M550129-03 Rev K
10.
Enter the actual output current of the second calibration point:
CAL:OUTP:CURR:FIVEPOINT 2
11.
Program the output of the third calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 32000
12.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 3.
13.
Enter the actual output current of the third calibration point:
CAL:OUTP:CURR:FIVEPOINT 3
14.
Program the output of the fourth calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 45000
15.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 4.
16.
Enter the actual output current of the fourth calibration point:
CAL:OUTP:CURR:FIVEPOINT 4
17.
Program the output of the fifth calibration point by sending the following
command string from the computer:
CAL:OUTP:CURR:DAC 62250
18.
Let the output settle and measure the output current with the current shunt
and the high precision voltmeter; this is value 5.
19.
Enter the actual output current of the fifth calibration point:
CAL:OUTP:CURR:FIVEPOINT 5
20.
To review entered data for five–point current calibration, issue the following
query:
CAL:OUTP:CURR:FIVEPOINT?
21.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL:UNLOCK “6867”
22.
Program the supply to store the calibration values in non-volatile memory:
CAL:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.