Calibration

SG Series Programming

7-8

M550129-03 Rev K

10.

Enter the actual output current of the second calibration point:

CAL:OUTP:CURR:FIVEPOINT 2

11.

Program the output of the third calibration point by sending the following

command string from the computer:

CAL:OUTP:CURR:DAC 32000

12.

Let the output settle and measure the output current with the current shunt

and the high precision voltmeter; this is value 3.

13.

Enter the actual output current of the third calibration point:

CAL:OUTP:CURR:FIVEPOINT 3

14.

Program the output of the fourth calibration point by sending the following

command string from the computer:

CAL:OUTP:CURR:DAC 45000

15.

Let the output settle and measure the output current with the current shunt

and the high precision voltmeter; this is value 4.

16.

Enter the actual output current of the fourth calibration point:

CAL:OUTP:CURR:FIVEPOINT 4

17.

Program the output of the fifth calibration point by sending the following

command string from the computer:

CAL:OUTP:CURR:DAC 62250

18.

Let the output settle and measure the output current with the current shunt

and the high precision voltmeter; this is value 5.

19.

Enter the actual output current of the fifth calibration point:

CAL:OUTP:CURR:FIVEPOINT 5

20.

To review entered data for five–point current calibration, issue the following

query:

CAL:OUTP:CURR:FIVEPOINT?

21.

Program the supply to unlock the non-volatile memory for calibration value

storage:

CAL:UNLOCK “6867”

22.

Program the supply to store the calibration values in non-volatile memory:

CAL:STORE

Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.