Error Code Description
201
Unexpected warm boot
This error means that the DIA processor experienced a warm boot that
was unexpected, and it may indicate an internal crash of the DIA
processor.
102
Incompatible unit type
This error is not used. It cannot occur.
100
Incompatibility error
This error is not used. It cannot occur.
0
No error
The error queue is empty.
-102
Syntax error
An unrecognized command or data type was encountered.
-108
Parameter not allowed
More arguments than expected were received.
-151
Invalid string data
Incorrect password. Manufacturer, model, or serial number string was
more than 16 characters. Invalid mnemonic.
-161
Invalid block data
The expected number of data values was not received.
-200
Execution error
An error/event number in the range [-299,-200] indicates that an error has
been detected by the instruments execution control block. The occurrence
of any error in this class shall cause the execution error bit (bit 4) in the
Event Status Register to be set. An execution error can be the result of:
•
A
in a 33 volt device.
•
A command could not be executed due to the current condition of the
device.
-203
Command protected
Attempted to store calibration values to EEPROM without unlocking.
-221
Settings conflict
Attempted to set output greater than soft limits or to set soft limits less than
output.
-222
Data out of range
Parameter exceeded range of valid values.
-225
Out of memory
There is not enough memory to perform the requested operation.
-241
Hardware missing
A legal command or query could not be executed because of a hardware
fault.
-284
Program currently running
A legal command or query could not be executed because a function is
currently running.
-292
Referenced name does not exist
-293
Referenced name already exists
-316
Checksum error