SG Series Programming

IEEE/RS232 Configuration and Remote Programming

M550129-03 Rev K

3-3

3.2

RS232 SETUP PROCEDURE

This section provides a quick reference for the configuration requirements for
RS232. Refer to Sections 3.3.1 and 3.3.2 for detailed information on the rear
panel configuration switches.

Parameter

Setting

Notes

Baud Rate

SGA models: Fixed, 19200

The baud rate for SGI models is

selectable through the front panel

using the Remote Menu; refer to

Section 3.8.11 of the SGI Operation

Manual, M550221-01.

SGI models: Selectable,

2400 to 19200

Data Bits

8

Stop Bits

1

Parity

None

Flow Control

None

Incoming Termination

Character

CR (Carriage Return):

HEX, 0x0d (DEC, 13)

Outgoing Termination

Character(s)

CR LF (Carriage Return and

Line Feed):

HEX, 0x0d 0x0a (DEC, 13 10)

Selectable using SCPI command,

SYST:NET:TERM; refer to

Section 6.9.2.


1. Build an RS232 communications cable per the pinout description

illustrated in Figure 3-4 (with crossover of signals Rx/Tx and CTS/RTS):

RJ-type 6P6C Plug

Terminal

D-Subminiature 9-Pin Connector

Female Socket

1

2

2

7

3

No Connection

4

5

5

3

6

8

Figure 3-4. RS232 Communications Cable Pinout