Description

Sorensen Power Supply SGx

IP Address

DHCP-acquired (Primary default*) If DHCP absent,
assigned via Auto-IP (Secondary default*)

IP Addressing mode

DHCP-acquired (Primary default*)

Subnet Mask

DHCP-acquired (Primary default*) If DHCP absent,
assigned via Auto-IP (Secondary default*)

Gateway

0.0.0.0

DNS Server

0.0.0.0

Listening Port

9221

User ID

admin

Password

password

Ping Echo

On

* Primary/Secondary defaults:

The Ethernet interface provides the opportunity to set both a Primary and
a Secondary IP configuration in the IP Configuration page (Section 5.5.2).
If the Primary fails, the system defaults to the Secondary configuration.
However, both setting “DHCP-acquired” and selecting “Auto IP Enabled”
together in the Primary configuration, prevents the power supply from
trying the Secondary configuration. Please see “TCP/IP Configuration”
and “Auto IP Enabled” under IP Configuration, Section 5.5.2 for more
detail.

4.3.3

Programming Resolution

SGA

SGI

Voltage

0.002% of full scale

0.002% of full scale

Current

0.002% of full scale

0.002% of full scale

Overvoltage
Protection

0.002% of full scale (full scale is 110% of
max output voltage.)

0.002% of full scale (full
scale is 110% of max
output voltage.)

4.3.4

Programming Accuracy

SGA

SGI

Voltage

± (0.1% of maximum output voltage)

± (0.1% + 0.1% of full scale)

Current

± (0.25% of full scale output current

± (0.1% + 0.4% of full scale)

Overvoltage
Protection

± (0.5% of max output voltage)

±

(0.5% + 0.5% of full scale)

(full scale 110% of max. output
voltage)

4.3.5

Readback Resolution

SGA

SGI

Voltage

0.002% of full scale

0.02% of full scale

Current

0.002% of full scale

0.02% of full scale