Calibration

SG Series Programming

7-4

M550129-03 Rev K

13.

Enter the actual output voltage value of the third calibration point:

CAL:OUTP:VOLT:FIVEPOINT 3

14.

Program the output of the fourth calibration point by sending the

following command string from the computer:

CAL:OUTP:VOLT:DAC 45000

15. Let the output settle and measure the voltage with a high precision

voltmeter; this is value 4.

16.

Enter the actual output voltage value of the fourth calibration point:

CAL:OUTP:VOLT:FIVEPOINT 4

17.

Program the output of the fifth calibration point by sending the following

command string from the computer:

CAL:OUTP:VOLT:DAC 62250

18. Let the output settle and measure the voltage with a high precision

voltmeter; this is value 5.

19.

Enter the actual output voltage value of the fifth calibration point:

CAL:OUTP:VOLT:FIVEPOINT 5

20.

To review entered data for five–point voltage calibration, issue the

following query:

CAL:OUTP:VOLT:FIVEPOINT?

21.

Program the supply to unlock the non-volatile memory for calibration

value storage:

CAL:UNLOCK “6867”

22.

Program the supply to store the calibration values in non-volatile

memory:

CAL:STORE

Allow 15 seconds for the non-volatile memory to be updated with
the new calibration values.

23.

Program the supply to lock the non-volatile memory for calibration value

protection.

CAL:LOCK