Calibration

SG Series Programming

7-10

M550129-03 Rev K

9.

Reset the output current to 0 amps.

CAL:OUTP:CURR:DAC 0

10.

Program the Digital Interface Adapter (DIA) to calculate the current readback

calibration gain and offset values:

CAL:OUTP:CURR:CALC

CAL:MEAS:CURR:CALC

11.

Program the DIA to unlock the non-volatile memory for calibration value

storage:

CAL:UNLOCK “6867”

12.

Program the Digital Interface Adapter (DIA) to store the date:

GPIB

:

CAL:DATE YYYYMMDD

Ethernet:

CAL:MOD:LASTCALDATE MM DD YYYY

CAL:MOD:NEXTCALDATE MM DD YYYY

13.

Program the DIA to store the calibration values in non-volatile memory:

CAL:STORE

Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.

14.

Program the DIA to lock the non-volatile memory for calibration value

protection

CAL:LOCK

Calibration is complete.