4.2.10 CALIBRATION Menus

Measurement Calibration:

Output Calibration:

The measurement calibration menu can be used to perform routine calibration of the internal
measurement system. The recommended calibration interval is 12 months. To enter the
calibration screens, the calibration password must be entered first.

The output calibration menu can be used to perform routine calibration of the voltage output and
programmable impedance. The recommended calibration interval is 12 months. To enter the
calibration screens, the calibration password must be entered first.

Note: Refer to chapter 6 for details on routine calibration procedures and equipment
requirements. Do not attempt calibration without consulting the user manual.

This menu also contains the LCD viewing angle adjustment.

ENTRY

DESCRIPTION

VIEW ANGLE

-10 to +10

LCD viewing angle adjustment.

CAL PWORD

Calibration password required to access all
calibration screens. The calibration password
is 5000. The password can be entered using
the keypad or shuttle followed by the ENTER
key.

Measurement Calibration Screens

MVOLT FS

Calibration coefficient for full-scale voltage
measurement.

MCURR FS

Calibration coefficient for full-scale current
measurement.