beautypg.com

AMETEK MX15 Series User Manual

Page 163

background image

User Manual - Rev M

California Instruments

MX15

163

104V for 1 minute.

122V for 1 minute.

115V for 1 minute.

The

key (backspace) will terminate the test at any time.

FREQUENCY

This test will change the output frequency in the following sequence:

360Hz for 1 minute.

440Hz for 1 minute.

400Hz for 1 minute.

The

key (backspace) will terminate the test at any time.

9.4.7.1

ABNORMAL TEST

From the MIL704 main menu (Figure 9-11) scroll to the ABNORMAL AC entry using the up and
down cursor keys. Press the ENTER key to select the ABNORMAL screens. The screen will
appear as shown in Figure 9-14.

Figure 9-14: Abnormal Screens

The ABNORMAL SCREEN has the following tests:

1. OVER VOLTAGE

2. UNDER VOLTAGE

3. OVER FREQUENCY

4. UNDER FREQUENCY

The above test can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.

OVER VOLTAGE

This test will change the output voltage for the selected phase in the following sequence:

For 400Hz and VFREQ:

180V for 50msec.

The voltage gradually decays with time to 125 volt by the following equation:
V = 124.6 + 2.77/t. For 0.05

t

6.925

Stay at 125V for 93 seconds before returning to 115V.

For 60Hz only:

180V for 3.34msec

The Voltage gradually decays with time to 122 volt by the following equation:
V = 121.7 + 0.583/t. For 0.00334

t

1.947