180

Number

Message String

Cause

Remedy

not be recalled.

overwrite old content.

6

"Limit memory lost"

Hardware configuration
settings lost.

Contact CI service department
at

[email protected]

to

obtain instructions on restoring
configuration data.

7

"System memory lost"

Memory corrupted.

Recycle power.

8

"Calibration memory
lost"

Calibration data lost.

Contact CI service department
at

[email protected]

to

obtain instructions on restoring
calibration data or recalibrate
unit.

9

"Start angle must be
first sequence"

Start phase angle in wrong
place..

Start phase angles can only
programmed at the start of a
transient list. Once a transient is
in progress, phase angle cannot
be changed.

10

"Illegal for DC"

Operation not possible in
DC mode.

Switch to AC or AC+DC mode.

11

"Duplicate sequence"

Transient list sequence
number already used.

User new or available sequence
number instead.

12

"Too many sequence"

Number of transient list
steps exceeds maximum.

Reduce the number of steps in
the transient list. (Max = 32 for
Series I or 100 for Series II).

13

"Missing list
parameter"

One or more transient list
parameters missing.

Check programmed lists.

14

"Voltage peak error "

Peak voltage exceeds
internal bus voltage

This error may occur when
selecting user defined wave
shapes with higher crest
factors. Reduce programmed
RMS value.

15

"Slew time exceed
dwell"

Time needed to slew to
final value is less than
dwell time.

Check dwell times in transient
list settings. Increase dwell time
or change slew rate for affected
parameter.

16

"Illegal during
transient"

Operation requested not
available while transient is
running.

Wait till transient execution is
completed or abort transient
execution first.

17

"Output relay must be
closed"

Operation not possible with
open relay

Close relay before attempting
operation. E.g. transient
execution requires output relay
to be closed.

18

"Trans. duration less
then 1msec"

Dwell time below minimum
or 1 msec

Increase dwell time to at least 1
msec.

19

"Clock and sync must
be internal"

Operation not possible with
external clock

Switch to internal sync.
(Default)

20

"Input buffer full"

Too much data received.

Break up data in smaller blocks.

21

"EOS Fault"

Hardware error reported by
EOS option

Cycle power on EOS to reset
error. If error persists, contact
CI service at

[email protected]

for

repair.

22

"Waveform harmonics
limit"

Harmonic contents of user
defined wave shape are

Reduce harmonic content or
reduce fundamental frequency