AMETEK MX15 Series User Manual
Page 87
![background image](https://www.manualsdir.com/files/771676/content/doc087.png)
User Manual
– Rev M
California Instruments
MX15
87
4.2.9
CONFIGURATION Menu
The configuration menu may be used to configure various aspects of the instrument such as the
serial port (including USB and LAN), IEEE-488/GPIB address and the power-on settings of the
supply.
ENTRY
DESCRIPTION
ADDRESS
0 - 31
Sets the selected IEEE / GPIB bus address for
the optional IEEE/GPIB interface.
Factory default is address 1. The shuttle knob
or the keypad can be used to set a value from
0 through 31. Do not use address 0 as this
address is typically reserved for the GPIB
controller.
BAUD RATE
9600
19200
38400
57600
115200
230400
460800
Sets the baud rate for the RS232
communications port. Factory default is
38400 baud. Available settings are 9600
through 115200 baud.
The same setting is used for USB and LAN
modes. For use with either USB or LAN, the
baud rate in this screen must be set to
460800.
Note: The shuttle knob can be used to scroll
through these selections.
PON STATE
REG0
RST
Determined power on state. This setting
selects either non-volatile REG0 to be recalled
automatically at power-on or factory default
(RST). Factory default is RST, which recalls
the factory settings. See below.
Factory default settings are:
Output relay
Open
Voltage Range
Low
Mode
AC
Voltage
0 Vrms
Frequency
60 Hz
Note that to use REG0 for power-on default,
the contents of the register must be
programmed first. See section 4.2.8. If an
empty register is selected, the power source
will revert back to RST (factory setting).