4.2.4

PROGRAM Menus

Figure 4-4: PROGRAM Menus

The PROGRAM menu is shown in Figure 4-4. It can be reached in one of two ways:

1. By selecting the MENU key, selecting the PROGRAM entry and pressing the Enter key.

2. By pressing the SET key.

The PROGRAM menu is used to change primary output parameters. Less frequently used
parameters are located in the CONTROL menu.

The following choices are available in the PROGRAM menus:

ENTRY

DESCRIPTION

VOLT

Programs the output voltage in Vrms. The voltage can be
changed from 0 to its max range value as determined by the
configuration settings and the selected voltage range by using
the keypad + Enter or the shuttle (if the voltage field is selected).

FREQ

Programs the output frequency. The frequency can be changed
from its min to its max value as determined by the configuration
settings by using the keypad + Enter or the shuttle (if the
frequency field is selected).

VRANGE

Selects 150V, 300V, or 400V voltage range (if available). The
value of this field can be changed with the shuttle as long as the
active pointer () points to the VRANGE entry.

PHASE

Selects the phase angle between the external clock and the
output of the AC source. If the clock source is internal, this
parameter has no effect.

FUNC

Selects the waveform for the selected phase. On MX15-1Pi
models, available choices are SINUSOID, SQUARE and
CLIPPED or any user defined waveform that was downloaded to
the AC source waveform memory using the RS232 or IEEE-488
interface.
This field is fixed to SINUSOID on MX15-1 models.

CLIP LEVEL

Sets the clip level for the CLIPPED sine wave in percent VTHD.