9.3.1.6

EMERGENCY TEST

From the DO160 MENU scroll to the EMERGENCY AC entry using the up and down cursor keys.
Press the ENTER key to select the EMERGENCY screens. The screen will appear as shown in
Figure 9-8.

Figure 9-8: Emergency Screens

The EMERGENCY SCREEN has the following tests:

1

VOLT FREQ MIN

2

VOLT FREQ MAX

3

VOLT UNBALANCE

The above tests can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.

VOLT FREQ MIN

Standard/Group

RTCA

A(CF)

A(NF)

A(WF)

Voltage

100

100

100

100

N/A

N/A

N/A

N/A

Frequency

360

360

360

360

Standard/Group

Group1

Group2

Group3

Voltage

104

104

104

N/A

N/A

N/A

Frequency

360

360

360

Table 9-7: Emergency Voltage and Frequency Minimum

Standard/Group

RTCA

Group1

Group2

Group3

Voltage

122

122

122

122

N/A

N/A

N/A

N/A

Frequency

440

440

650

800

Standard/Group

Group1

Group2

Group3

Voltage

122

122

122

N/A

N/A

N/A

Frequency

440

650

800

Table 9-8: Emergency Voltage and Frequency Maximum

This test will set the voltage and frequency to a level defined by Table 9-7. The test will last for 30
minutes. The test will be repeated using the voltage from Table 9-8 and frequency from Table
9-7. T
he

key (backspace) will terminate the test at any time.