9.8 Option

–411: IEC 61000-4-11 Voltage Dips and Interruptions

9.8.1

General

The IEC 61000-4-11 option is capable of performing IEC 61000-4 section 11 voltage dips, short
interruptions and voltage variations immunity tests. On MX15 Series AC sources, the user can
only perform single phase tests.

Operation of the

9.8.2

Front Panel Entry

No front panel operation is available for the IEC 61000-4-13 option on the MX15 Series AC
power source. Instead, use of the provided MXGUI Windows software is required to use this
option.
The -411 option can be programmed using SCPI bus commands. For details, refer to the MX
Series programming manual P/N 7003-961.

9.8.3

Standard Revisions and EUT Classes

The

–411 option supports both the first (1994-06) and the second edition (2004-03) of the IEC

61000-4-11 test standard.

Generic tests files are distributed with the MXGui program for both editions of the test standard.
Files applicable to Edition 2.0 have ED20 in their file name. Do not mix these files, as the data
setup will not be correct if you do. To load a test file, select the Mode (Dips or Vars) and test
standard revision first, then use the File, Open menu to load the test parameters. Test
parameters can be a function of the EUT class. The different files provided with the program
cover the various EUT classes. The relevant EUT class 1, 2, 3 or X is listed in the file names.

Refer to the MXGUI on line help for information on using the IEC411 test screen..