Number

Message String

Cause

Remedy

operation.

-311

"Memory error"

Waveform memory
checksum error.

May be the result of incomplete
user-defined waveform
download. Check interface and
try downloading waveform
again. Successful download
may clear this error condition.

Alternatively, use
TRAC:DEL:ALL command to
clear waveform memory.

-314

"Save/recall memory
lost"

User setup register
contents lost

Store setup in same register
again.

-315

"Configuration memory
lost"

Hardware configuration
settings lost.

Contact CI service department
at

[email protected]

to

obtain instructions on restoring
configuration data.

-330

"Self-test failed"

Internal error

Contact CI service department
at

[email protected]

-350

"Queue overflow"

Message queue full.

Too many messages. Read
status using SYST:ERR query
until 0, "No Error" is received
indicating queue empty.

-400

"Query error"

Unable to complete query.

Check programming manual for
correct query format and
parameters

-410

"Query
INTERRUPTED"

Query issued but response
not read.

Check application program for
correct flow. Response must be
read after each query to avoid
this error.

-420

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

-430

"Query
DEADLOCKED"

Query cannot be
completed

Check application program for
multiple queries

-440

"Query
UNTERMINATED"

Query incomplete.

Check for terminator after query
command.

1

"Output volt fault"

Output voltage does not
match programmed value.

Load exceeds current limit and
unit is in Constant Voltage (CV)
mode of operation. Reduce load
or increase CL setting.
Output voltage is driven above
programmed voltage by
external influence (Load,
voltage kickback, etc.)

2

"Current limit fault"

Current limit exceeded.

Load exceeds current limit and
unit is in Constant Voltage (CV)
mode of operation. Reduce load
or increase CL setting

3

"Temperature fault"

Temperature of heat sink
too high.

Reduce load. Ensure proper
airflow and exhaust clearance.
Check fan(s) for operation.

4

"External sync. error"

Could not sync to external
sync signal.

External sync signal missing,
disconnected or out of range.

5

"Initial memory lost"

Power on settings could

Save power on settings again to