beautypg.com

AMETEK MX15 Series User Manual

Page 159

background image

User Manual - Rev M

California Instruments

MX15

159

9.4.6

Steady State AC Tests

Scroll to the NORMAL ST MENU entry using the up and down cursor keys. Press the ENTER
key to select the STEADY STATE screens. The screen will appear as shown in Figure 9-12

Figure 9-12: Steady State Menu

The MIL704 Steady state screens have the following tests:

1. VOLTAGE

2. FREQUENCY

3. VOLT UNBALANCE

4. PHASE DIFFERENCE

5. VOLT MODULATION

6. FREQ MODULATION

7. VOLT TRANSIENT

8. FREQ TRANSIENT

9. DISTORTION

The above tests can be selected by scrolling to the highlighted selection using the up and down
cursor keys and the ENTER key to start the selected test.

VOLTAGE

This test will change the output voltage in the sequence shown in Table 9-12.

SEQUENCE

VOLTAGE

TIME

400Hz/VFREQ

60Hz only

1

108

110

1 minute

2

118

125

1 minute

3

115

115

1 minute

Table 9-12: Steady state voltage

The

key (backspace) will terminate the test at any time.

FREQUENCY

This test will change the output frequency in the sequence shown in Table 9-13.

SEQUENCE

FREQUENCY

TIME