AMETEK MX15 Series User Manual
Page 160
User Manual
– Rev M
California Instruments
MX15
160
400Hz
VFREQ
60 Hz
1
393
360
59
1 minute
2
407
800
61
1 minute
3
400
SSF
60
1 minute
Table 9-13: Steady state frequency
The
key (backspace) will terminate the test at any time.
VOLT UNBALANCE
This test will change the output voltage for the selected phase only in the following sequence:
112V for 1 minute.
118V for 1 minute.
115V for 1 minute.
The test will be repeated on three phase systems to include all three phases if the coupling is set to all.
The
key (backspace) will terminate the test at any time.
PHASE DIFFERENCE
This test applies to three phase systems only. The phase angle for the selected phase will
change relative to phase A in the following sequence:
If phase B is selected:
236
for 1 minute.
244
for 1 minute.
240
for 1 minute.
If phase C is selected:
116
for 1 minute.
124
for 1 minute.
120
for 1 minute
VOLTAGE MODULATION
This test will vary the output voltage by
2.5V rms over a period of one second. The test will last
for 2 minutes. The
key (backspace) will terminate the test at any time.
FREQUENCY MODULATION
REVISION
D
E
F (400Hz /VFREQ)
F (60HZ)
MODULATION
7Hz
4Hz
4Hz
0.5Hz
Table 9-14: Frequency Modulation
This test will vary the output frequency as defined by Table 9-14 over a period of one minute.
The test will last for 4 minutes. The
key (backspace) will terminate the test at any time.