6.2 Front Panel Calibration Screens

The calibration screens for output or measurement calibration can be selected from the OUTP
CAL and MEAS CAL screens.

To select the OUTPUT CALIBRATION screen press ENTER on the OUTP CAL field. This will
bring up the CAL PWORD screen. To prevent unauthorized access to calibration data, a
password must be entered to access any calibration screen. The calibration password is always
"5000" and may be entered using the numeric keypad. Once entered, the calibration screens
remain accessible until the MX unit is powered down.

Type 5000 and press the ENTER key to show the OUTPUT CALIBRATION screen.

To select the MEASUREMENT CALIBRATION screen, follow the same steps as outlined above
but select the MEAS CAL entry instead of OUTP CAL. If another CALIBRATION screen has
been accessed since power-up, no password is needed. Otherwise, enter the same password
as indicated above.