AMETEK MX15 Series User Manual
Page 18
User Manual
– Rev M
California Instruments
MX15
18
Parameter
Specification
Frequency: step, sag, surge, sweep
Voltage and Frequency: step, sweep
IEEE-488 Interface:
SH1, AH1, T6, L3, SR1, RL2, DC1, DT1
IEEE 488.2 and SCPI
Response time is 10 ms (typical)
RS232C Interface:
Bi-directional serial interface
9 pin D-shell connector
Handshake:
CTS, RTS
Data bits:
7, 8
Stop bits:
1,2
Baud rate:
9600 to 115,200 bps
Syntax: IEEE 488.2 and SCPI
USB Interface:
Standard USB 2.0 peripheral.
Data transfer rate: 460,800 bps
Syntax: IEEE 488.2 and SCP.
Note:
Use of the USB port to control more than one power source
from a single PC is not recommended, as communication may not be
reliable. Use GPIB interface for multiple power source control.
Note: Not available on older MX15 models.
LAN Interface:
Option
–LAN. When the LAN interface is installed, the RS232 interface is
disabled.
RJ45 Connector, 10BaseT, 100BaseT or 1000BaseT,
Data transfer rate: 460,800 bps
Protocol: TCP/IP.
Syntax: IEEE 488.2 and SCP
Note: Disconnect any USB connection when using the LAN interface.
Current Limit Modes:
Two selectable modes of operation:
1. Constant current mode (voltage folds back with automatic recovery)
2. Constant voltage mode with trip-off (Relays open).