VOLT FREQ MAX

This test will set the voltage and frequency to a level defined by Table 9-8. The test will last for 30
minutes. The test will be repeated using the voltage from Table 9-7 and frequency from Table
9-8. T
he

key (backspace) will terminate the test at any time.

VOLT UNBALANCE

This test is not available on the MX15 Series since only a single phase output is available.

9.3.1.7

ABNORMAL TEST

From the DO160 MENU Scroll to the ABNORMAL AC entry using the up and down cursor keys.
Press the ENTER key to select the ABNORMAL screens. The screen will appear as shown in
Figure 9-9.

Figure 9-9: Abnormal Screen

The ABNORMAL SCREEN has the following tests:

1

VOLT MAX

2

VOLT MIN

3

VOLT UNBALNCE

4

VOLT SURG

5

VOLT DROP

6

FREQ TRANSIENTS

The above test can be selected by scrolling to the highlighted selection using the up and down
key and the ENTER key to start the selected test.

VOLT MAX

Standard/Group

RTCA

Group1/A(CF)

Group2/A(NF)

Group3/A(WF)

Voltage

1

97

97

104/100

97

97

3

N/A

N/A

N/A

N/A

N/A

Frequency

400

400

370

360

360

Table 9-9: Abnormal Voltage Minimum

Standard/Group

RTCA

Group1/ACF)

Group2/A(NF)

Group3/A(WF)

Voltage

1

134

134

122

134

134

3

N/A

N/A

N/A

N/A

N/A

Frequency

400

400

430

650

800

Table 9-10: Abnormal Voltage Maximum