AMETEK MX15 Series User Manual

Page 125

background image

User Manual - Rev M

California Instruments

MX15

125

6.3.1

Measurement Cal

AC Volt Full-scale:

Program the output to the 300 VAC range. Close the output
relay. Program the output to 240 VAC and 60 Hz. Go to the
MEASUREMENT CALIBRATION screen. Enter the actual AC
output voltage for the VOLT FS parameter and press the
ENTER key. Save this value by pressing the SET key.

Note: This process may take as long as a few minutes to
complete after the enter key is pressed.

AC Current Full-scale:

Apply a load to the output. Refer to Table 6-1. Program the
output to 120 VAC on the 150 VAC range and 60 Hz.
Observe the actual output current and enter this value for the
CURR FS parameter. Press the ENTER key. Save this value
by pressing the SET key.

DC Volt Full-scale:

Program the output to the 400 VDC range. Close the output
relay. Program the output to 320 VDC. Go to the
MEASUREMENT CALIBRATION screen. Enter the actual AC
output voltage for the VOLT FS parameter and press the
ENTER key. Save this value by pressing the SET key.

Note: This process may take as long as a few minutes to
complete after the enter key is pressed.

DC Current Full-scale:

Apply a load to the output. Refer to Table 6-1. Program the
output to 160 VDC on the 200 VDC range.
Observe the actual output current and enter this value for the
CURR FS parameter. Press the ENTER key. Save this value
by pressing the SET key.

6.3.2

Measurement Calibration Summary

The following Table is a summary of the preceding calibration steps. The value indicated by the
External DVM is called V

AC

or V

DC

. The current measured by the current shunt is called I

AC

or

I

DC

.

TITLE

PROGRAM/LOAD PARAMETERS

PARAMETER

ADJUST TO

AC MODE

AC Volt Full-
scale

300 VAC Range, 240 VAC, 60 Hz, no
load

VOLT FS

V

AC

AC Current Full-
scale

150 VAC Range, 120 VAC, 60 Hz, full
load to 90% of max current range.

CURR FS

I

AC

DC MODE

DC Volt Full-
scale

400 VDC Range, 320 VDC, no load

VOLT FS

V

AC

DC Current Full-
scale

200 VDC Range, 160 VDC, full load to
90% of max current range.

CURR FS

I

AC

Table 6-2: Measurement Calibration Table