beautypg.com

AMETEK MX15 Series User Manual

Page 78

background image

User Manual

– Rev M

California Instruments

MX15

78

4.2.5

CONTROL Menus

Figure 4-5: CONTROL Menus

The CONTROL menu is shown in Figure 4-5 and can be reached by selecting the Menu key,
selecting the CONTROL entry using the DOWN cursor key and then pressing the Enter key.

The CONTROL menu is used to change secondary output parameters. The following choices are
available in the CONTROL menus:

ENTRY

DESCRIPTION

SENSE

Selects internal or external (remote) voltage sense mode. If INT is
selected, the voltage is sensed at the output terminal block. If EXT is
selected, the voltage is sensed at the external sense connector. If
external sense is selected, care must be taken to connect the
external sense lines at the load. For sense leads longer than 1
meter, twisted pairs should be used.

SYNC

Selects the external sync mode if available. Default is internal sync,
which means a free running time base. The time base can be
synchronized to an external sync signal by selecting external sync
mode.

CLOCK

Selects internal or external clock source. The MX Series controller
uses an open-air crystal time base with an accuracy of 100 ppm. The
external clock mode is used to support the

–LKS option. For use as

an auxiliary unit in a clock and lock system, this field must be set to
EXT. A unit with

–LKS option can be used stand-alone if needed by

setting the INT clock mode.

INT

Default, internal clock.

EXT

Auxiliary unit (-LKS) driven by master (-LKM) clock
input.
Note: When selecting EXT mode, make sure the
Clock and Lock BNC cables are connected to the
Master (-LKM) unit. If not, there will be no output on
the

–LKS unit. See section 3.10 for connection

information.