9.9 Option

–413: IEC 61000-4-13 Interharmonics Test

9.9.1

General

The IEC413 option is capable of performing IEC 61000-4 section 13 Harmonics and inter
harmonics low frequency immunity tests. The tests are based on IEC 61000-4-13:2002-03, First
Edition. It is assumed that the user has a copy of the test standard available.

9.9.2

Initial Setup

The user must set the operating voltage and close the output relay prior to the start of test. The
following set of parameters must be set before the start of test.

1. Frequency to 50 or 60 Hz.

2. Voltage mode to AC.

3. Waveform to sine wave.

9.9.3

Front Panel Entry

No front panel operation is available for the IEC 61000-4-13 option on the MX15 Series AC
power source. Instead, use of the provided MXGUI Windows software is required to use this
option.
The -413 option can be programmed using SCPI bus commands. For details, refer to the MX
Series programming manual P/N 7003-961.,

9.9.4

Tests Performed

The IEC1000-4-13 test consists of several types of tests. These tests can be run individually or
in sequence (ALL). The following tests are available:

1. Harmonic combination test flat curve and over swing.

2. Sweep in frequency and resonance frequency detection.

3. Individual harmonics and inter harmonics.

4. Meister curve test.