Error Messages
- Setup software TM-H1 TM-Navigator User’s Manual -
Error Messages
This section explains information about the displayed error messages that appear when an error occurs.
Item
Display contents
Remedy
Reference
page
File operations Failed to read the file ( ).
Verify that the correct file is set.
* The path name is entered in the
parentheses ( ).
Check whether the selected file is an TM-
Navigator program file (*.prs) or an
environment setting file (*.ens).
Failed to read the environment settings.
( )
* The failed item is entered in the parentheses ( ).
Check the unreadable environment setting
items and set correctly on the controller.
Master image is short. Change short master
image for non-reg.
This occurs when the master image saved
during program saving does not exist in the
same folder as for the selected file.
Put the master image (.bmp) in the same
folder.
Sending and
receiving
settings
Communication failed.
Set the controller to "Measurement screen
(during measurement)".
-
Transfer setting failed.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement screen
(during measurement)".
Receive settings failed.
• Check whether communication is possible
between the controller and computer.
• Set the controller to "Measurement screen
(during measurement)".
Head setting
Image capture failed.
Perform "Trigger input" with the controller to
capture the image.
The threshold exceeds the setting range. Set
the value between 10 and 99.
Set the threshold within the displayed range.
Master
registration
The master image is not registered and
cannot be set. Register the master image in
"Master Image registration".
Register the master image.
Correction angle is over the setting range.
Set the correction angle to within ±20
q.
Position
correction
The correlation value exceeds the setting
range.
Set the correlation value within the setting
range.
Measurement
setting
The master image is not registered and
cannot be set. Register the master image in
"Master Image registration".
This is displayed when the master image is
not registered when "Search" is selected in
"Measurement type".