4-14

- Setup software TM-H1 TM-Navigator User’s Manual -

Data Storage

Displa

ying

the

Measurem

ent Data

(4) [Excel transfer start] button

Stored data can be transferred to Excel and saved as a file.

For operation method, see "Transferring the storage data to Excel" (page 4-18).

For the contents of the Excel file, see "Excel File" (page A-5).

(5) [Data storage] button

Displays the [Data storage] dialog box.

This dialog box can be used to configure the controller's data storage settings.

(6) Amount of stored data and current status display

Displays the amount of data that is stored on the controller and current operation status of the controller.

(7) Graph display information for each OUT setting

Changes the graph display for each OUT setting.

(8) Cursor information display area

Displays the information for cursor A and cursor B on the graph display area.

Cursors A and B can be moved by dragging the mouse.

(9) [Read storage data] button

Reads the data that is stored in the connected controller on TM-Navigator.

No. of data points

Sets the number of data points to store.

Setting range: 1 to 65536

Skipping

Sets the number of data point skipped before being stored as data.

Setting range: 1 / 2 / 5 / 10 / 20 / 50 / 100

Received

Receives all of the storage settings (program No. 00 to 15 and environment settings) from the

controller.

Send

Sends all of the storage settings (program No. 00 to 15 and environment settings) that are

being edited from TM-Navigator to the controller.

Close

Closes the [Data storage] dialog box.

Switches between showing and hiding the graph.

Check this checkbox to display the graph.

Displays the name of the graph. The name can be changed.

Moves the graph in the horizontal direction.

Displays the color used for the graph line. Any color can be set.