Vesuv 3.0

6. Database

• 89

Show variables and formulas

Menu item "View, Show variables+formulas"

The variables window will be displayed. You need this win-
dow for recalculation of results with modified variables or
formulas.

The column headers of this window mean the following:

No

Running number.

Sub-determination

Number of sub-determination, important

only for 726 Titroprocessor.

Name

Name of variable.
In the formulas you may find variable names
which are different from these names for Titri-
nos and KF-Coulometer and for 726 Titroproc-
essor.

Property

of the variable.

Value

of the variable

E

Indicates if this value can be edited.
0: Value cannot be edited.
1: Value can be edited.

M

Indicates if this value has been modified.
0: Value has not been modified.
1: Value has been modified.

List of calculation variables for Titrinos and 756
KF Coulometer

The names of the variables in the "Variables + formulas" ta-
ble is different from the symbols used in the formula for the
following variables:

Symbol in formula

Name

C00

SmplSize

C21

Id1

C22

Id2

C23

Id3