5. Datalogger
• 45
. Day of week, e.g. each Monday or each 2nd
Monday
. Day of month, e.g. every first day of the month
When you alter the time interval the time
counter is reset to zero.
Time of day
Enter the time when the message is to appear.
The
Calculate first expiration date from
today
: From the current date. If you select this
option the next date will be calculated immedi-
ately. This can be seen in the date list in the
lower part of the window.
last date
: From the last date set. If the corre-
sponding test has never been monitored then
the date of the first start of the Vesuv Datalog-
ger will be used.
last log file entry
: From the deadline on which
the test was last carried out, i.e. the last date to
have been protocolled for this test. If the test log
file is still empty then the date of the first start of
the Vesuv Datalogger will be used.
Fixed date
Select "Fixed date" when the monitoring is to respond on a
particular day, e.g. on 1 May 1999.
Date
Set the date. The
rent date.
Time of day
Enter the time when the message should ap-
pear. The
Message repeating frequency
Repeat message each
Enter the frequency with which the message
should appear. The message will be repeated
until monitoring has been confirmed and/or a
new date has been set.
This setting applies to both incremental moni-
toring as well as monitoring to a fixed date.
Information about dates
Last log file entry
of the currently activated test (calibration,
GLP test or service). If the test log file is still
empty then the date of the first start of the Ve-
suv Datalogger will be used.