beautypg.com

Metrohm VESUV User Manual

Page 74

background image

Vesuv 3.0

6. Database

• 69

Step by step procedure

1.

Select a data set, which contains the desired item, and
go to menu item "View, Properties".

2.

Select an item you don't need from the list. This item will
be overwritten by the new selection.

3.

Press button . A window with the following table
appears:

Title

Enter the column header.

Table

Select the table of the database, which contains
the item.
The table "Vars" will usually contain the desired
item. Values which are evaluated by the analytical
device (such as EP, rates, PK/HNP etc.) are in the
table "PointRes".
The other table contain information about the de-
vice and the method.

Sub-determination

Set 1.

For 726 Titroprocessor set the corresponding sub-
determination (1...5).
If you need information about the device (Device
type etc.), set 0.

Name

Select the name of the item.

The list contains all items, which are present in the
selected data set and in the table you selected
above. If you don't find the desired item, try to
change the table.

Property

Select the property of the item.

Usually you have to select "Value" (also for
strings) or possibly "Mean" (if you wish to repre-
sent the mean which has been calculated by your
analytical device).

Type

Select the data type (String/Number).
This selection may change the sort type.

Statistics

If you selected type "Number", you may acti-

vate statistics calculation for this column ( ).

4.

Quit the window with and represent the column
( ).

5.

Repeat steps 2-4 for each new item.

If you quit the window "Determinations, Properties" with
, the list will be sorted alphabetically