52

• 5. Datalogger

Vesuv 3.0

5.

Use the button to store the meth-
ods in the file.
This procedure may take some time.

The standard directory for the methods is set under the menu
item "Options, Settings, Default directories".

Presentation options

Select methods

After they have been read in all the meth-

ods are marked automatically. This option is
used for e.g. method backup.

Show open directories

After they have been read in all the directories
are shown open, so that the individual methods
can be directly marked for storing.

Restore methods

Menu item:

"Device, Restore methods"

Mouse:

right mouse in device

Methods which are stored in a PC can be copied into a con-
nected device. Methods from Titrinos, the 756 KF Coulome-
ter and 730 Sample Changer can be restored. For Sample
Changers the methods and the configuration (with the posi-
tion tables of the special racks) are restored.

If the menu item is gray the instrument must first be stopped

(

or no methods from the instrument can be restored.

Before you restore the methods, make sure that

the device is in its inactive basic state

no data transfer is taking place

the instrument has no RS error (e.g. error 42)

the device window is stopped (i.e. "Not ready, off line:

button

)

the option "Vesuv can send data to device" has been ac-
tivated under "Properties, Communication"

Note: before the methods are restored Vesuv will stop the
analytical device. Any waiting reports will be lost!