Vesuv 3.0

2. Connecting devices

• 17

707, 768 KF Oven

Cables

PC 25pin

6.2125.060

PC 9pin

6.2125.110

If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit and no parity. The handshake
must be set as follows:

HWs (if the Oven is connected to a COM port)

SWchar (if the Oven is connected via a Softswitch)

The RS parameters can be set on the Oven in key
, >RS232 settings.

Notes

If the oven is connected to the PC, then transmit at least
the automatic report at the end of the titration. You can
also send additional reports which can be stored in the
database and/or printed out.

If the oven is connected to an instrument via an RS cable
then do not send an oven report.