Vesuv 3.0

7. Troubleshooting

• 101

Problem

Remedy

Missing characters in raw
reports

Check the correct setting of the FIFO buffer:

Win 95, 98: , Settings, Control panel,
System, Device manager, Ports, COMX, Port
settings, Advanced: Set transmit and receive
buffer to "low". Don't switch it off!

Win NT: , Settings, Control panel, Ports,
COMX, Settings, Advanced: Don't activate FIFO
buffer.

No data reception

Check the RS settings on your device and the
port parameter in the datalogger.

Check the cable: Do you use the correct cable?
Is the cable correctly plugged in (the end with
"RS232C" at the device, the end with "IBM PC"
at the PC side)?

Does your device send a report at the end of the
determination?
Is this report on COM1 of the analytical device (if
you have more than one)?

If you work with a Softswitch, check if the inputs
are switched over (LED is changing from input to
input). If inputs are not switched over:
. Set switch of Softswitch at the outermost posi-
tion in the direction of the PC.
. SOFT1.X is the Softswitch connected to COM1
of the PC, etc.
. Connect Softswitch to another COM of the PC.

Switch your analytical devices off/on.

Restart your PC.

For pH Meters and Conductometers: Send only
reports without included curves (short reports).

For 691 pH Meter: Set Handshake of pH Meter
to HS=0.

726 Titroprocessor:
No data reception anymore

If data reception is not ok after switching on the
Titroprocessor:
Initialize the RS interface at the Titroprocessor in
"Config, Interfaces".