beautypg.com

Re-evaluation, recalculation – Metrohm VESUV User Manual

Page 92

background image

Vesuv 3.0

6. Database

• 87

Re-evaluation, recalculation

Re-evaluation, recalculation

Right mouse button on the curve, Re-evaluation

The curves can be evaluated graphically.

Step by step procedure

1.

Select the type of evaluation.

2.

By using the keys (and ) and the left
mouse button you can move the selected line along the
curve with the cursor. The EP-coordinates are continu-
ously displayed in the "Result" field.
Position the line(s).

3.

Use the button to add the new EP. This
manually evaluated EP receives a different symbol. The
symbol can be specified under Properties, Endpoints.

4.

The evaluation lines can be retained and printed out
("Properties, Endpoints, Show evaluation lines").

5.

The original condition is returned to with the menu item
"Data, Original results" of the database.

Deleting an EP

1.

Select the EP: Click onto the text "EPX". The current EP
is shown in the "Active endpoint" field.

2.

Delete the EP with the button.

Evaluation possibilities

Move the cursor on the curve with the key and
the left mouse button.

Evaluation with intersection lines. The tangents are ap-
plied to the curve. The EP is defined as the intersection
of the tangents. This evaluation is used, for example, in
conductometric and photometric titrations.
Move Tangent 1 along the curve with the key
and the left mouse button; move Tangent 2 with the
key and the left mouse button.
The curve sections can be zoomed (mouse).
For fine positioning, curve smoothing can be switched
on under Properties.