Device manager – Metrohm VESUV User Manual

Page 41

background image

36

• 5. Datalogger

Vesuv 3.0

Device manager

Menu item "Device, Device manager".

This menu item is only active when all device windows are

stopped (button

).

Configure instrument

If you are connecting devices for the first time:

1.

See pages 8ff for cable connections and extended PC
interfaces.
Open a configuration: Menu item "File, Configuration,
Open" and give your configuration a name.

If you already have connected instruments:

1.

Stop all device windows (button

).

Further:

2.

Select the menu item "Device, Device manager".

3.

Add a new instrument with the button

.

A balance is selected with the button

.

4.

A list of device types appears. Give the device a name
and select the device type from the list.

5.

The dialog for properties of the device window appears.
From the sheet "Communication" select the port to which
your instrument is connected.
A balance must be connected directly to a COM port or
via an extension board (not via the Softswitch).

6.

Select the database in which the determinations with this
device are to be stored, sheet "Database".
Also check the database template.
The database is created when the window "Device man-
ager" is closed.

7.

Exit from the dialog and test the connection to your in-

strument: press the button

and carry out the

test>.
If the button is inactive transmit a report from
your instrument to Vesuv.

8.

Make the necessary settings on the device, see pages
8ff.