• 7. Troubleshooting

Vesuv 3.0

Problem

Remedy

726 Titroprocessor:
Device window remains in
status "Data reception"

The interface of the Titroprocessor, which is
connected to the PC, can't be printer interface at
the same time (Setting at the Titroprocessor in
Config, Interfaces).

A curve was sent by the Titroprocessor. The end
of the report could not be recognized.

Stop the Titroprocessor

and restart it

.

After switching on the Titroprocessor it may have
sent garbled characters.

Stop the Titroprocessor

and restart it

.

See also device settings of 726 Titroprocessor,
page 9.

Balances

Problem

Remedy

No balance data in the bal-
ance window

Check the RS settings on your balance and the
port parameter in the datalogger.

The balance can't be connected to a Softswitch.
Connect it directly to a PC COM or to a Moxa or
USB board resp

Switch balance off/on.

Restart your PC.

Do you use the correct cable and is the cable
correctly plugged in ?

No balance data in the de-
vice. They are correct in the
balance window

Did you assign the balance to this device (see
status bar of device window)?

In the settings of the device, the menu item
"Device window, Properties, Communication,
Vesuv can send data to device" has to be active

in the datalogger.

With automatic assignment of balance: Did you
set the device for automatic balance assignment
also?

With automatic assignment of balance: Is your
analytical device in the sample size request?

Check if the correct balance type is set in your
analytical device (key , >peripheral
units, for 726 Titroprocessor "Config, Inter-
faces").

Did you select the correct balance type in the
datalogger?

Remove the 6.2125.010 adapter from the RS
interface of your analytical device.