Connecting devices – Metrohm VESUV User Manual

Page 13

background image

8

• 2. Connecting devices

Vesuv 3.0

Connecting devices

Titrino

Cables

Titrino 25pin, PC 25pin

6.2125.060

Titrino 25pin, PC 9pin

6.2125.110

Titrino 9pin, PC 25pin

6.2125.110

Titrino 9pin, PC 9pin

6.2134.040

If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit and no parity. The handshake
must be set as follows:

HWs (if the Titrino is connected to a COM port)

SWchar (if the Titrino is connected via a Softswitch)

The RS parameters can be set on the Titrino in key
, >RS settings.

Notes

At the end of the titration at least the following reports
should be transmitted:
Full;Calc;MpList (if present)
In the mode CAL: Calib.
Set on the Titrino under ", >Report" (COM1).
You can also send additional reports which can be stored
in the database and/or printed out.
Important: Don't send any curves from the Titrino to Ve-
suv! The Vesuv curve is created from the measuring
point list which the Titrino transmits.

If you are working with TIP and require a separate record
for TIP in the database, then at the end of the TIP se-
quence a pause should be made which is longer than the
"Timeout" of Vesuv.

Titrinos with 2 RS interfaces

e.g. 751, 758, 784, 785 Titrinos: Connect the PC to COM
port 1 of the Titrino.

751, 758 Titrinos

If you use Italian for the instrument dialog then you re-
quire the program versions 5.751.0011 and 5.758.0011
or higher.

701 KF Titrino
Transmit the full report. Setting on the Titrino under
", >preselections"