Connecting devices – Metrohm VESUV User Manual
Page 13
8
• 2. Connecting devices
Vesuv 3.0
Connecting devices
Titrino
Cables
Titrino 25pin, PC 25pin
6.2125.060
Titrino 25pin, PC 9pin
6.2125.110
Titrino 9pin, PC 25pin
6.2125.110
Titrino 9pin, PC 9pin
6.2134.040
If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit and no parity. The handshake
must be set as follows:
•
HWs (if the Titrino is connected to a COM port)
•
SWchar (if the Titrino is connected via a Softswitch)
The RS parameters can be set on the Titrino in key
Notes
•
At the end of the titration at least the following reports
should be transmitted:
Full;Calc;MpList (if present)
In the mode CAL: Calib.
Set on the Titrino under "
You can also send additional reports which can be stored
in the database and/or printed out.
Important: Don't send any curves from the Titrino to Ve-
suv! The Vesuv curve is created from the measuring
point list which the Titrino transmits.
•
If you are working with TIP and require a separate record
for TIP in the database, then at the end of the TIP se-
quence a pause should be made which is longer than the
"Timeout" of Vesuv.
•
Titrinos with 2 RS interfaces
e.g. 751, 758, 784, 785 Titrinos: Connect the PC to COM
port 1 of the Titrino.
•
751, 758 Titrinos
If you use Italian for the instrument dialog then you re-
quire the program versions 5.751.0011 and 5.758.0011
or higher.
•
701 KF Titrino
Transmit the full report. Setting on the Titrino under
"