6 Devices

IC Net 2.0

86

Manual injection

Sample vial number

Number of sample vial to be used for man-

ual injection.

Selection of the injection number (

1...4

).

Needle lowering type

Injection number

Position where the needle is moved to in

case of multiple injections (

1...4

).

Absolute position

Position where the needle is moved to in

case of single injections (

1...4

).

Rinse type

Standard shift

Use

previous

or

next

vial for rinsing.

Nonstandard shift

For every sample, skip this number of vials

and use this vial for rinsing.

No rinse between different sample vials

No rinsing.

Data acquisition started by software

If this option is enabled, the data acquisi-

tion is started with "Inject". No cable is re-

quired between the injection valve and the

interface.

If this option is disabled, the 6.2115.070

cable is needed to start data acquisition.

Start injection.

For detailed information about the 750 Autosampler see 750 In-
structions for Use
.

Links

The

Links

tab of the

750 Autosampler

is used for COM port selec-

tion and settings (details see section 5.2.4).