Metrohm solvent delivery unit, Metrohm sdu features, Metrohm sdu icon – Metrohm IC Net 2.0 User Manual
Page 77: Metrohm sdu window, 6 metrohm solvent delivery unit

6.6 Metrohm solvent delivery unit
IC Net 2.0
69
Advanced
The
Advanced
tab of the
709 IC Pump
window is only available for a
connected system.
Correction factor
Flow correction factor to adjust the dis-
played flow to the actual flow measured
with the aid of a measuring cylinder.
Entry range:
0.90 ... 1.10
Read current
Correction factor
value.
Send current
Correction factor
value to the
709 IC Pump. This parameter is not stored
in the system file (
*.smt
) as long as the file
is not saved.
6.6
Metrohm solvent delivery unit
6.6.1
Metrohm SDU features
A Metrohm SDU (Solvent Delivery Unit) system can be used for
high-pressure gradient mixture of one to four eluents. For each
eluent, a 709 IC Pump is used. Each 709 IC Pump must be con-
nected to the 762 IC Interface using a 6.2134.080 or 6.2134.090
cable. Examples for interconnections with the 709 IC Pump can
be found in the 762 Instructions for Use.
The outlet capillaries of the 709 IC Pumps must be connected to
a mixing chamber from where one inlet capillary leads to the in-
jection valve of the 733 IC Separation Center.
For detailed information about the 709 IC Pump see 709 Instruc-
tions for Use.