Metrohm IC Net 2.0 User Manual
Page 45
5.2 762 IC Interface
IC Net 2.0
37
Event command (2
nd
column)
Setting the event output lines 1
…
7.
Selection:
0
(line off, inactive, open)
1
(line on, active, 0 V)
p
(pulse, pulse length 150 ms)
*
(leave line in current status)
For entry of the first value, enter
1
,
0
,
p
or
*
.
For entry of the other values, move the cur-
sor in front of the value to be changed and
enter
1
,
0
,
p
or
*
.
Add new remote command.
Delete selected remote command.
Activate the defined remote commands for
insertion into the time program.
Events setup
The
Events setup
tab of the
762 System [#]
window contains the
startup values for the seven event output lines of system 1 or
system 2 of the 762 IC Interface. These startup values are sent
each time the system is connected or a determination is started.
Each event output line can be set to the following values:
0
Line off, inactive (open)
1
Line on, active (0 V)
For technical information see 762 Instructions for Use.