beautypg.com

771 compact interface, 771 compact interface features, 771 icon on the toolbar – Metrohm IC Net 2.0 User Manual

Page 46: 771 compact interface window, 3 771 compact interface

background image

5 Interfaces

IC Net 2.0

38

5.3

771 Compact Interface

5.3.1

771 Compact Interface features

The Metrohm 771 Compact Interface is a two channel Analog-

to-Digital Converter to acquire two analog signals from two differ-

ent detectors simultaneously or time independent. It contains the

following parts:

Two 24-bit Analog-to-Digital Converters (often abbreviated

as ADC) to acquire two analog signals from two different de-

tectors simultaneously (for example conductivity detection and

UV detection).

Two external start inputs.
The 771 Compact Interface is normally connected to a PC COM

port using the 6.2134.100 connecting cable. The 6.2128.170 ca-

ble is used to acquire the analog signal from the 732 IC Detector

(or another detector) and to start the data acquisition with contact

closure from the 733 IC Separation Center.
For detailed information about the 771 Compact Interface see
771 Instructions for Use.

5.3.2

771 icon on the toolbar

The

771 icon

is available on the toolbar if a 771 Compact Inter-

face has been installed with the

New system wizard

or by using the

option of the

WORKPLACE

window (see section

5.1.2). By clicking this icon the

771 Compact Interface

window for

parameter settings opens.

5.3.3

771 Compact Interface window

771 icon / Open

The

771 Compact Interface

window for parameter settings is

opened by selecting this menu option with the right mouse button

or by clicking the

771 icon

on the toolbar. It consists of three tabs

Init

,

RS settings

, and

Links

.

Init

The

Init

tab of the

771 Compact Interface

window contains data ac-

quisition parameters for the 771 Compact Interface.

Label

Optional label to name the interface with

maximum 8 characters.

ROM version

Instrument program version number.

Measuring per second

Number of data points measured per s.

Entry range:

10, 20, 30, 50, 60 points/s