Metrohm IC Net 2.0 User Manual
Page 83

6.7 752 Pump Unit
IC Net 2.0
75
Manual
The
Manual
tab of the
752 Pump Unit
window for manual operation
is only available for a connected system.
Pump
Start pump drive.
Stop pump drive.
Start pump with startup hardware
Automatic start of pump drive with
Startup
hardware
.
Time program
On the
Program
tab of the
752 Pump Unit
window a user-defined
time program can be entered. This program is started automati-
cally as defined in the
Start mode
window (see section 4.4.3) ei-
ther at the moment the determination is started (
Start with determi-
nation
) or at the moment the sample is injected (
Start with inject
).
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.