beautypg.com

Peak editor, Switching on/off the peak editor, Peak editor functions – Metrohm IC Net 2.0 User Manual

Page 163: Moving the cursor, 4 peak editor

background image

8.4 Peak editor

IC Net 2.0

155

8.4

Peak editor

8.4.1

Switching on/off the peak editor

IC NET / Process / Peak editor

The peak editor mode is used for subsequent manual correction

of the automatic integration for chromatogram peaks (see section
7.4
). It can be used to select the most important points for the

peak evaluation (start, end and top of peak, valley between

peaks) and move them to the required position.
If the peak editor mode is enabled, the vertical bar cursor ap-

pears in the chromatogram window. At the same time, the

Peak

menu appears in the menu bar and the peak editor icons appear

in the icon bar of the main window.

The peak editor mode is switched on or off by clicking on

or

by pressing [ Alt ] + [ C ]. It is also possible to click the right

mouse button somewhere in the chromatogram window and to

select the

Peak editor

item.

You cannot edit the peak pattern while the component table is

active and vice versa, when the peak editor mode is switched on

the component table option is disabled.

8.4.2

Peak editor functions

The functions of the peak editor can be triggered with the corre-

sponding menu items of the

Peak

menu, by the peak editor sym-

bols in the symbol bar or with key combinations.

IC NET / Peak / Undo

Undo the last action.

IC NET / Peak / Insert peak

[ Insert ]

Add a peak to the chromatogram.

IC NET / Peak / Delete peak

[ Delete ]

Delete selected peak.

IC NET / Peak / Select nearest point

Move the cursor to the nearest start, top, end, or valley point and

select the peak.

IC NET / Peak / Select start point

Move the cursor to the beginning of the nearest peak and select

the peak.

IC NET / Peak / Select top point

Move the cursor to the top of the nearest peak and select the

peak.