Sample queue control, Sample queue overview table, Start sample queue – Metrohm IC Net 2.0 User Manual
Page 172: Pause sample queue, Cancel last run, Reset sample queue, 2 sample queue control

9 Sample queue
IC Net 2.0
164
9.2
Sample queue control
9.2.1
Sample queue overview table
IC NET / File / Open / Sample queue
SYSTEM / System / Sample queue
These two menu items open the
OPEN
window for opening an
existing sample queue file
*.que
or creation of a new sample
queue file
*.que
by entering a new name. After confirmation with
the sample queue overview table is opened showing the
current status of the sample queue:
No
Row number.
System
System file to be used for the determina-
tion.
Ident
User defined identifier for chromatogram.
Will be placed into appropriate passport
field when starting the chromatogram.
Vial
Autosampler vial position to take sample
from.
Inj-s
Number of injections for the same vial po-
sition (only 1 injection is permissible for
Level
≥≥ 1
).
Started
Indication whether the sample has been
started or not:
0
sample not started
1
sample started
2
…
…n
number of injections done
Level
Calibration level (see section 7.5) for the
sample: Level
0
stays for normal analysis,
levels
1
and greater for calibration runs.
Correct filling of this column enables to cal-
culate calibration coefficients automatically.
Volume
Injected volume of sample in
µ
L.