Metrohm IC Net 2.0 User Manual
Page 69
6.3 732 IC Detector
IC Net 2.0
61
The
Program
tab contains the two following subpages:
Program
Main time program with all program
steps.
Remote configuration
Possibility for creation of user-defined
remote commands.
Program
On the
Program
subpage, program steps including time, program
instruction and parameter can be entered.
First column
Time at which program instruction is ap-
plied.
Entry range:
0.0 ... 999.9 min
If no time is entered, the program instruc-
tion is applied together with the last in-
struction with time entry.
Second column
Program instruction (see below).
In addition to these predefined instructions,
user-defined remote commands can be
entered if activated on the
Remote configura-
tion
tab.
Third column
Parameter for program instruction (see
below).
ENABLED
Enable program start (a disabled program
is not started).
Will be executed by
Selection if the time program will be exe-
cuted by the
PC
or the
732 unit
.
Add new program instruction.
Delete selected program instruction.
Test the time program (error messages are
displayed if program is wrong).